USER MANUALCONTINUOUS INTERMITTENTMODEL SERIES: PM3300 (shown) PM3400SAVE THESE INSTRUCTIONSCAUTIONFederal (USA) law restricts this device to sale by
9CONTINUOUS INTERMITTENTDISPOSAL INSTRUCTIONSDispose of the Vacuum Regulator in accordance with the local regulations.Please RecycleWARNINGProduct sho
10CONTINUOUS INTERMITTENTLIMITED WARRANTY AND LIMITATION OF LIABILITYPrecision Medical, Inc. warrants that the Medical Vacuum Regulator (the Product)
DECLARATION OF CONFORMITYPrecision Meidcal, Inc300 Held DriveNorthampton PA 18067Emergo Europe (European Ofce)Motenstraat 152513 BH, The HagueThe Net
1CONTINUOUS INTERMITTENTCONTENTSRECEIVING / INSPECTION ... 2INTENDED USE ...
2CONTINUOUS INTERMITTENTRECEIVING / INSPECTIONRemove the Precision Medical, Inc. Vacuum Regulator from the packaging and inspect for damage. If there
3CONTINUOUS INTERMITTENTSPECIFICATIONSGAUGE RANGE:PM3300: 0 - 200 mmHg - Full Vacuum*PM3300E: 0 - 200 mmHg (0 - 26 kPa) *PM3300EHV: 0 - 300 mmHg (0 -
4CONTINUOUS INTERMITTENTOPERATING INSTRUCTIONSCAUTIONInspect the Vacuum Regulator for visual damage before use, DO NOT USE if damaged.NOTE: • Overow
5CONTINUOUS INTERMITTENTPARTS DESCRIPTIONCAUTIONMissing or illegible labels must be replaced, contact Precision Medical, Inc.PARTS LISTNo. Description
6CONTINUOUS INTERMITTENTREPAIR KITSAnalog Part# Digital Part#PM3300 / PM3300D Vac Reg RK6300 RK6300DPM3300HV / PM3300DHV Vac Reg RK6300HV RK6300DHVPM3
7CONTINUOUS INTERMITTENTCAUTION• DO NOT autoclave or immerse in liquid. This will cause damage to the Vacuum Regulator and will void the warranty.•
8CONTINUOUS INTERMITTENTCLEANING / DECONTAMINATION(As needed)1. Attach a working Vacuum Regulator with a continuous regulated mode to a minimum vacuu