USER MANUAL300 Held Drive Tel: (+001) 610-262-6090Northampton, PA 18067 USA Fax: (+001) 610-262-6080ISO 13485 Certied www.precisionmedical.comCAU
91. Position the cylinder so the oxygen cylinder valve outlet is pointing away from the user and any other person (s).2. Before installing the Conserv
103. Slowly open cylinder valve counterclockwise, until completely open.4. Turn dial on the Conserver until indicating pointer is aligned with the p
11MAINTENANCE / CLEANING1. Disconnect all connections before cleaning.2. After each use, clean exterior of the Conserver with a cloth dampened with mi
12TROUBLESHOOTINGIf the Conserver fails to function, consult the Troubleshooting Guide below. If problem cannot be solved, consult your Provider.REPLA
13ACCESSORIESDESCRIPTION PART #Carry Bag M6 503920Carry Bag M4/M6 504184Carry Bag ML6/M9 504185Carry Bag with shoulder strapModel 195405NG ShownConser
14LIMITED WARRANTY AND LIMITATION OF LIABILITYPrecision Medical, Inc. warrants that the EasyPulse5 Oxygen Conserving Regulator (the Product) will be f
DECLARATION OF CONFORMITY Precision Medical, Inc. 300 Held Drive Northampton, PA 18067, USA Emergo Europe (European Ofce) Molenstraat 15 251
1CONTENTSRECEIVING / INSPECTION ...2INTENDED USE ...
2RECEIVING / INSPECTIONRemove the Precision Medical, Inc. EasyPulse5 Oxygen Conserving Regulator (Conserver) from the packaging and inspect for damage
3SAFETY INFORMATION - WARNINGS AND CAUTIONSDANGERIndicates an imminently hazardous situation which, if not avoided, will result in death or serious in
4WARNING• ALWAYS conrm prescribed dose before administering to patient and monitor on a frequent basis.• Always follow standards for Medical Gas Pr
5CAUTION• Only personnel instructed and trained in its use should operate the Conserver.• The Conserver contains magnetic, ferrous material that may
6TECHNISCHE DATENInlet Pressure Range: 300 - 3000 psi (2068 - 20684 kPa)Pressure Gauge Accuracy: 3 - 2 - 3% of full scaleDimensions: (Are approximate
7DIAGRAMS / COMPONENT DESCRIPTIONCAUTIONMissing or illegible labels must be replaced, contact Precision Medical, Inc. Inlet Seal(O-ring)DialCannulaCon
8INSTALLATIONWARNING• Read this User Manual before installing or operating the Oxygen Conserving Regulator.• Inspect the Conserver and cylinder valv