Oxygen Concentrator
Service Manual
Model PM4300 Series
300 Held Drive
Northampton Pa 18067 (USA)
Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-353-1240
ISO 13485 Certified
www.precisionmedical.com
Page view 0
1 2 3 4 5 6 ... 21 22

Summary of Contents

Page 1 - Service Manual

Oxygen Concentrator Service Manual Model PM4300 Series 300 Held Drive Northampton Pa 18067 (USA) Tel: (+001) 610-262-6090 Fax: (+001) 610-262-608

Page 2 - Contents

8 Miscellaneous 506639 – Nylon Hose Clamp (Flowmeter & O2 Sensor) 505932 – 11/16” Pinch Clamp (Sieve Bed Hose) 506763 – Tubing Sleeve Clamp

Page 3 - Preface

9 Inlet Filter: 1. Remove the Inlet Filter Cover by inserting the tip of a flat blade screwdriver into the slot shown and prying off. 2. Grasp ce

Page 4 - CAUTION

10 Visual Indicator / Problem Audible Alarm Probable Cause Remedy RED General Alarm Indicator flashes once, device momentarily stops functioning, an

Page 5 - Theory of Operation

11 Visual Indicator / Problem Audible Alarm Probable Cause Remedy Limited or no oxygen flow. None Dirty or obstructed bubble humidifier, or leak pre

Page 6

12 Front Panel Assembly Replacement (PM4350 – 507593 PM4351 – 507321) Disassembly Tools and equipment required: #2 Phillips Screwdriver 5/16

Page 7 - Specifications:

13 Reassembly (Reference Disassembly Photos) 1. Lay Front Panel Assembly in front of device. 2. Connect all Wires as shown in step 14 of the Fron

Page 8 - Component Description

14 Screws (X4) PCB Insulating Paper Front Panel Dial Flowmeter Replacement (506620) Disassembly Tools and equipment required: # Phillips Screwdri

Page 9 - Internal

15 Compressor Assembly Replacement (507322) Disassembly Tools and equipment required: #2 Phillips Screwdriver Diagonal Cutting Pliers Needle Nos

Page 10 - WARNING

16 Reassembly (Reference Disassembly Photos) 1. Place Compressor Assembly, center port facing to the back of the device, onto the Compressor Tray

Page 11 - Troubleshooting

17 Sieve Bed Assembly Replacement (506418) Disassembly Tools and equipment required: #2 Phillips Screwdriver ¼” Hex Nut Driver Diagonal Cutting Pl

Page 12

Contents Preface ...

Page 13 - Alarm Indicators

18 Final Test Test Requirements for Stationary Oxygen Concentrator 1. Ensure Device Power Switch is in off position “O”. 2. Plug Device into a 1

Page 15 - Reassembly

300 Held Drive Northampton Pa 18067 (USA) Tel: (+001) 610-262-6090 Fax: (+001) 610-262-6080 Toll Free Tel: 800-272-7285 Toll Free F

Page 16

1 Preface Precision Medical, Inc. 300 Held Drive Northampton, PA 18067 USA Customer Service / Tech Support Phone: 1-610-262-6090 Phone: 1-

Page 17

2 Safety Symbols information - Warnings and Cautions DANGER Indicates an imminently hazardous situation which, if not avoided, will result in dea

Page 18

3 Theory of Operation Air enters the system through the compressor intake muffler/filter (1). The compressor (2) increases the air pressure to

Page 19

4 Warnings and Cautions WARNING Oxygen supplied from this device is for supplemental use and is not intended to be life supporting or life sustai

Page 20 - Final Test

5 Specifications: Weights: Device: 29.6 lbs (13.4 kg) Device Boxed for shipping: 33.9 lbs (15.38 kg) Dimensions: Height: 17.0 in (43.2 cm)

Page 21

6 Component Description External Part # Description Includes 506829 Base & Power Cord Assembly 506333 – Base 505933 – Power Cord 507321 PM43

Page 22 - Northampton Pa 18067 (USA)

7 Internal Part # Description Includes 507323 Compressor Cover Assy 505429 – Fan 505595 – Fan Wire Harness 507001 – Cover